ASTM F750 - 26

Standard Practice for Evaluating Acute Systemic Toxicity of Material Extracts by Systemic Injection in the Mouse

Aviso: Las normas ASTM en lectura se visualizan a través del área de cliente. Los archivos no se pueden descargar
Fecha edición: 2026-02-01
En Vigor
Idiomas disponibles: Inglés
Resumen:

Keywords: acute toxicity tests; biocompatibility; intravenous injection; intraperitoneal injections; mouse/mice; test animals;
Scope:

1.1 This practice covers methods used in an acute toxicity test for detecting systemic toxicity associated with leachables from materials used in medical devices.

1.2 The liquids injected into the mouse are those obtained by Practice F619 where the extraction vehicles are saline, vegetable oil, or other liquids simulating human body fluids.

1.3 Two procedures are outlined: Method A for intravenous injection and Method B for intraperitoneal injection.

1.4 This practice is one of several developed for the assessment of the biocompatibility of materials. Practice F748 may provide guidance for the selection of appropriate methods for testing materials for a specific application.

1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.7 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ICS: 11.220 - Medicina veterinaria
CTN: F04.16 - F04.16

Anulaciones Normas

Anula a F750 - 20

Otras Relaciones

Es referenciada por F2148 - 25

Es referenciada por E3219 - 25

Es referenciada por F748 - 25

El libro en palabras del autor

Ultricies magna feugiat malesuada sociosqu varius vivamus cubilia parturient, himenaeos vitae vehicula nam placerat netus urna platea, nostra rutrum felis mattis penatibus velit quisque.

Button
Preguntas frecuentes ¿Tienes alguna duda sobre nuestros productos?

Respuesta 2

Desde la web

Libros y normas